Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Human IL-8 standard curve comparison data.
Standard curve comparison between human IL-8 SimpleStep ELISA® kit and traditional ELISA kit from leading competitor. SimpleStep ELISA® kit shows a 3-fold increase in sensitivity.

Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Example of human IL-8 standard curve in Sample Diluent NS.
The IL-8 standard curve was prepared as described. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Serum from ten individual healthy human female donors was measured in duplicate.
Interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean IL-8 concentration was determined to be 61.6 pg/mL with a range of 62 – 174.9 pg/mL; two individuals (Donor #3 and Donor #9) measured below the minimal detectable dose.

Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Example of human IL-8 standard curve.
Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Interpolated concentrations of native IL-8 in human serum.
The concentrations of IL-8 were measured in duplicates, interpolated from the IL-8 standard curves and corrected for sample dilution. Undiluted samples are as follows: serum 50%. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean IL-8 concentration was determined to be 63.2 pg/mL in serum.

Sandwich ELISA - Human IL-8 ELISA Kit (ab214030)
Interpolated concentrations of native IL-8 in human PBMC cell culture supernatant.
The concentrations of IL-8 were measured in duplicates, interpolated from the IL-8 standard curves and corrected for sample dilution. Undiluted samples are as follows: unstimulated 1:400 and stimulated 1:400. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). Measured values were interpolated from the IL-8 Standard Curve diluted in Sample Diluent NS and corrected for dilution factor. Mean of duplicate values +/-SD are graphed. The mean IL-8 concentration was determined to be 2.2 ng/mL in unstimulated, 119.7ng/mL in stimulated, and undetectable in media.